Operator’s Manual
Rev F 12/08/14 pg. 17
3.3 Remove the white Teflon Insert from the Extraction Vessel.
3.4 Remove the Bag Holder from the Teflon Insert.
3.5 Place up to 15 bags in the Bag Holder.
3.6 Place the Bag Holder in the Teflon Insert.
3.7 Press ENTER on the Keypad.
The Display should now show the screen to the right.
4.
Set the Extraction Time.
4.1 Using the arrow keys on the Keypad, select the Extraction time. (The recommended time is 60
minutes.)
4.2 Press ENTER on the Keypad.
The Display should now show the screen to the right.
5.
Set the Extraction Temperature.
5.1 Using the arrow keys on the Keypad, select the Temperature. (The recommended maximum
temperature is 90°C.)
5.2 Press ENTER on the Keypad.
The Display should now show the screen to the right.
6.
Insert solvent into the Extraction Vessel.
6.1 Ensure that the Extraction Vessel is cool.
6.2 Pour 350ml of solvent directly into the empty Extraction Vessel.
Insert Solvent
<ENTER>
Select temp 90
(^v)... <ENTER>
Select time 60
(^v)... <ENTER>
Teflon Insert
Bag Holder
Seitenansicht 16